Feasibility study of using a "travelling" CO<sub>2</sub> and CH<sub>4</sub> instrument to validate continuous in situ measurement stations
S. Hammer, G. Konrad, A. Vermeulen, O. Laurent, M. Delmotte, A. Jordan, L. Hazan, S. Conil, I. Levin

Feasibility study of using a "travelling" CO<sub>2</sub> and CH<sub>4</sub> instrument to validate continuous in situ measurement stations

Science & Mathematics Published 2013 Verification Pending

Overview & Synopsis

In the course of the ICOS (Integrated Carbon Observation System) Demonstration Experiment a feasibility study on the usefulness of a travelling comparison instrument (TCI) was conducted in order to evaluate continuous atmospheric CO2 and CH4 measurements at two European stations. The aim of the TCI is to independently measure ambient air in parallel to the standard station instrumentation, thus providing a comprehensive comparison that includes the sample intake system, the instrument itself as well as its calibration and data evaluation. Observed differences between the TCI and a gas chromatographic system, which acted as a reference for the TCI, were −0.02 ± 0.08 μmol mol−1 for CO2 and −0.3 ± 2.3 nmol mol−1 for CH4. Over a period of two weeks each, the continuous CO2 and CH4 measurements at two ICOS field stations, Cabauw (CBW), the Netherlands and Houdelaincourt (Observatoire Pérenne de l'Environnement, OPE), France, were compared to co-located TCI measurements. At Cabauw mean differences of 0.21 ± 0.06 μmol mol−1 for CO2 and 0.41 ± 0.50 nmol mol−1 for CH4 were found. For OPE the mean differences were 0.13 ± 0.07 μmol mol−1 for CO2 and 0.44 ± 0.36 nmol mol−1 for CH4. Offsets arising from differences in the working standard calibrations or leakages/contaminations in the drying systems are too small to explain the observed differences. Hence the most likely causes of these observed differences are leakages or contaminations in the intake lines and/or their flushing pumps. For the Cabauw instrument an additional error contribution originates from insufficient flushing of standard gases. Although the TCI is an extensive quality control approach it cannot replace other quality control systems. Thus, a comprehensive quality management strategy for atmospheric monitoring networks is proposed as well.

Advertisement
SPONSOR CONTENT
Indexed from repository: HAL Open Science (hal.science)
Loading document from verified source...
Keyboard Shortcuts: ← / → for pages • + / - for zoom • 0 to fit width Powered by Mozilla PDF.js v3.11

Document Information

Document Title Feasibility study of using a "travelling" CO<sub>2</sub> and CH<sub>4</sub> instrument to validate continuous in situ measurement stations
Author / Organization S. Hammer, G. Konrad, A. Vermeulen, O. Laurent, M. Delmotte, A. Jordan, L. Hazan, S. Conil, I. Levin
Subject & Discipline Multidisciplinary Academic Open Research
Categories Science & Mathematics
Publication Year 2013
License / Rights Open Access / Creative Commons
Resource Verification Listed PDF (Verification Pending)
Original Host Website HAL Open Science (hal.science)
File Format application/pdf (Portable Document Format)
Language EN
Catalog ID DOC-104753
Original Source Link Visit Source Page on hal.science »