Home »
Academic Articles »
Applied Physics »
A Novel Passive Implantable Differential Mechanism to Restore Individuated Finger Flexion during Grasping following Tendon Transfer Surgery: A Pilot Study
A Novel Passive Implantable Differential Mechanism to Restore Individuated Finger Flexion during Grasping following Tendon Transfer Surgery: A Pilot Study
Suraj Chakravarthi Raja
(Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, CA 90007, USA)
Won Suk You
(AstroForge, Inc., 15261 Connector Ln, Huntington Beach, CA 92649, USA)
Kian Jalaleddini
(iKinesia Inc., 490 Boul Provencher, Brossard, QC J4W 1Y4, Canada)
Justin C. Casebier
(Collaborative Robotics and Intelligent Systems (CoRIS) Institute, School of Mechanical, Industrial and Manufacturing Engineering (MIME), Oregon State University, Graf Hall 315, Corvallis, OR 97331, USA)
Nina R. Lightdale-Miric
(Children’s Hospital, Los Angeles, CA 90027, USA)
Vincent R. Hentz
(The Chase Center for Hand and Upper Limb Surgery, Stanford University, Stanford, CA 94305, USA)
Francisco J. Valero-Cuevas
(Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, CA 90007, USA)
Ravi Balasubramanian
(Collaborative Robotics and Intelligent Systems (CoRIS) Institute, School of Mechanical, Industrial and Manufacturing Engineering (MIME), Oregon State University, Graf Hall 315, Corvallis, OR 97331, USA)
Journal: Applied Sciences
•
• MDPI AG
• ISSN: 2076-3417
Abstract
Tendon transfer surgery is often used to restore hand grasp function following high median-ulnar nerve palsy. This surgery typically reroutes and sutures the tendon of the extensor carpi radialis longus (ECRL) muscle to all four flexor digitorum profundus (FDP) tendons of the hand, coupling them together. This makes it difficult to grasp irregularly shaped objects. We propose inserting a novel implantable passive device between the FDP tendons to surgically construct a differential mechanism, enabling the fingers to individually adapt to the irregular contours during grasping. These passive implants with no moving parts are fabricated from biocompatible materials. We tested the implants’ ability to create differential flexion between the index and middle fingers when actuated by a single muscle in two human cadaver hands using a computerized closed-loop control paradigm. In these cadaveric models, the implants enabled significantly more differential flexion between the index and middle fingers for a wide range of donor tendon tensions. The implants also redistributed fingertip forces between fingers. When grasping uneven objects, the difference in contact forces between fingers reduced by nearly 23% compared to the current suture-based surgery. These results suggest that self-adaptive grasp is possible in tendon transfers that drive multiple distal flexor tendons.
Suraj Chakravarthi Raja, Won Suk You, Kian Jalaleddini, et al. (2023). "A Novel Passive Implantable Differential Mechanism to Restore Individuated Finger Flexion during Grasping following Tendon Transfer Surgery: A Pilot Study". Applied Sciences. https://doi.org/10.3390/app13095804